Westgard multirule for calculated laboratory tests

Yükleniyor...
Küçük Resim

Tarih

2006

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Walter De Gruyter & Co

Erişim Hakkı

info:eu-repo/semantics/closedAccess

Özet

Background: In clinical laboratories, many test results such as low-density lipoprotein (LDL)-cholesterol can be calculated using the results of specifically measured tests. However, quality control programs deal only with measured tests and there is no adequate quality assessment procedure for calculated tests. I aimed to apply the Westgard multirule to calculated tests as a part of routine quality procedures. Methods: I prepared a Levey-Jennings chart for calculated tests and all their measured components (Equation chart) and then applied the Westgard multirule to these tests. Results: Applying the Westgard multirule to calculated LDL-cholesterol (and other calculated tests) was simple, and increased the reliability of the results of the calculated tests. Conclusions: Checking the reliability of only measured tests by quality control procedures before reporting patient results may be inadequate. Applying the Westgard multirule to calculated tests as part of total quality management will increase the reliability of test results.

Açıklama

Coskun, Abdurrahman/0000-0002-1273-0604
WOS: 000241193900003
PubMed: 17032128

Anahtar Kelimeler

laboratory management, quality control, reliability, Westgard multirule

Kaynak

Clinical Chemistry And Laboratory Medicine

WoS Q Değeri

Q2

Scopus Q Değeri

Q1

Cilt

44

Sayı

10

Künye