Westgard multirule for calculated laboratory tests
Yükleniyor...
Dosyalar
Tarih
2006
Yazarlar
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Walter De Gruyter & Co
Erişim Hakkı
info:eu-repo/semantics/closedAccess
Özet
Background: In clinical laboratories, many test results such as low-density lipoprotein (LDL)-cholesterol can be calculated using the results of specifically measured tests. However, quality control programs deal only with measured tests and there is no adequate quality assessment procedure for calculated tests. I aimed to apply the Westgard multirule to calculated tests as a part of routine quality procedures. Methods: I prepared a Levey-Jennings chart for calculated tests and all their measured components (Equation chart) and then applied the Westgard multirule to these tests. Results: Applying the Westgard multirule to calculated LDL-cholesterol (and other calculated tests) was simple, and increased the reliability of the results of the calculated tests. Conclusions: Checking the reliability of only measured tests by quality control procedures before reporting patient results may be inadequate. Applying the Westgard multirule to calculated tests as part of total quality management will increase the reliability of test results.
Açıklama
Coskun, Abdurrahman/0000-0002-1273-0604
WOS: 000241193900003
PubMed: 17032128
WOS: 000241193900003
PubMed: 17032128
Anahtar Kelimeler
laboratory management, quality control, reliability, Westgard multirule
Kaynak
Clinical Chemistry And Laboratory Medicine
WoS Q Değeri
Q2
Scopus Q Değeri
Q1
Cilt
44
Sayı
10