Comparison of central corneal thickness and intraocular pressure measured with two different tono/pachymeter devices in non-glaucomatouse children

dc.authoridAnkarali, Handan Camdeviren/0000-0002-3613-0523
dc.authoridTeberik, Kuddusi/0000-0003-3141-0531
dc.authorwosidAnkarali, Handan Camdeviren/P-1058-2016
dc.contributor.authorTeberik, Kuddusi
dc.contributor.authorEski, Mehmet Tahir
dc.contributor.authorAnkarali, Handan
dc.date.accessioned2021-12-01T18:49:10Z
dc.date.available2021-12-01T18:49:10Z
dc.date.issued2021
dc.department[Belirlenecek]en_US
dc.description.abstractPurpose: The goal of this research is to compare the intraocular pressure (IOP) and the mean central corneal thickness (CCT) values obtained from the measurements with Nidek NT-530P and Canon TX-20P devices with the values obtained by ultrasound pachymetry (UP) and Goldmann applanation tonometry (GAT) in children. Methods: This prospective study was conducted with 119 healthy children. The measurements were repeated three times for each eye. The intraclass correlation coefficient (ICC) was used to assess the correlation between the measurements obtained from different devices. The Bland-Altman plot was used to analyze the agreement between two different devices graphically. The measurements were taken in the same order in all subjects: TX-20P, NT-530P, UP, and GAT. Results: The mean age of the children was 10.1 +/- 3.2 (6-17) years. The mean CCT values for the eyes were 568.90 and 569.68 (TX-20P), 571.44 and 566.37 (NT-530P), 564.77 and 564.67 mu m UP, (right and left, respectively). The highest correspondence observed for CCT was between UP and NT-530P devices (ICC, 0.982, the left eye). The mean IOP measurements for Canon TX-20P, NT-530P, and GAT were 16.5 +/- 3.2, 16.3 +/- 3.2, and 16.8 +/- 3.6 mmHg for the right eyes, respectively, while those for the left eyes were 16.6 +/- 3.1, 16.2 +/- 3.3, and 16.8 +/- 3.7 mmHg. The highest correspondence was noted between measurements obtained using GAT and Nidek NT-530P devices (ICC, 0.945, in the left eye). Conclusion: Tono/pachymetry devices in the healthy children are provides significant agreements with the gold standard methods for IOP and CCT measurements. Clinical Trial Registration number: 2016-94en_US
dc.identifier.doi10.1177/1120672121991492
dc.identifier.endpage3017en_US
dc.identifier.issn1120-6721
dc.identifier.issn1724-6016
dc.identifier.issue6en_US
dc.identifier.scopus2-s2.0-85100592219en_US
dc.identifier.scopusqualityQ2en_US
dc.identifier.startpage3010en_US
dc.identifier.urihttps://doi.org/10.1177/1120672121991492
dc.identifier.urihttps://hdl.handle.net/20.500.12684/10676
dc.identifier.volume31en_US
dc.identifier.wosWOS:000677799200001en_US
dc.identifier.wosqualityQ4en_US
dc.indekslendigikaynakWeb of Scienceen_US
dc.indekslendigikaynakPubMeden_US
dc.indekslendigikaynakScopusen_US
dc.language.isoenen_US
dc.publisherSage Publications Ltden_US
dc.relation.ispartofEuropean Journal Of Ophthalmologyen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectCanon TX-20Pen_US
dc.subjectNidek NT-530Pen_US
dc.subjectintraocular pressureen_US
dc.subjectcentral corneal thicknessen_US
dc.subjectultrasound pachymeteren_US
dc.subjectApplanation Tonometryen_US
dc.subjectTono-Pachymeteren_US
dc.subjectUltrasounden_US
dc.subjectPentacamen_US
dc.subjectContouren_US
dc.subjectSystemen_US
dc.subjectErroren_US
dc.titleComparison of central corneal thickness and intraocular pressure measured with two different tono/pachymeter devices in non-glaucomatouse childrenen_US
dc.typeArticleen_US