Precision and accuracy studies with Kajaani fiber length analyzers

Yükleniyor...
Küçük Resim

Tarih

2007

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Asian Network for Scientific Information

Erişim Hakkı

info:eu-repo/semantics/openAccess

Özet

The aim of this study was to test the measurement precision and accuracy of the Kajaani FS-100 giving attention to possible machine error in the measurements. Fiber length of pine pulps produced using polysulfide, kraft, biokraft and soda methods were determined using both FS-100 and FiberLab automated fiber length analyzers. The measured length values were compared for both methods. The measurement precision and accuracy was tested by replicated measurements using rayon stable fibers. Measurements performed on pulp samples showed typical length distributions for both analyzers. Results obtained from Kajaani FS-100 and FiberLab showed a significant correlation. The shorter length measurement with FiberLab was found to be mainly due to the instrument calibration. The measurement repeatability tested for Kajaani FS-100 indicated that the measurements are precise. © 2007 Asian Network for Scientific Information.

Açıklama

Anahtar Kelimeler

Accuracy; Fiber length; Kajaani fiberlab; Kajaani FS-100; Precision