Performance of Cardiocheck (Tm) Portable Lipid Analyser

dc.contributor.authorBuğdaycı, Güler
dc.contributor.authorSezer, Sevilay
dc.contributor.authorKoca, Yüksel
dc.contributor.authorTurhan, Turan
dc.date.accessioned2024-12-28T20:19:40Z
dc.date.available2024-12-28T20:19:40Z
dc.date.issued2005
dc.description.abstractTo verify whether precision of lipids analyses by a new portable device, CardioCheck(TM), were in agreement with the guidelines of the National Cholesterol Education Program (NCEP). Serum samples from hundred outpatients were collected for the determination of total Cholesterol (TC), HDL-cholesterol (HDL-C) and triglycerides (TG). These samples were analysed simultaneously by the CardioCheck(TM), and by the automatic enzymatic methods routinely used at the biochemistry laboratory. Precision was determined by repeating 20 times the evaluation of normal and high control sera. Precision for panel TC was 3.29 % (NCEP ? 3 %), for panel HDL-C was 6.1 % (NCEP ? 6%) and for panel TG was 18.9 % (NCEP ? 5 %), TC was 5.3 % , HDL-C 14.5 % , TG was 15.7 %. Precision lipids evaluation by the CardioCheck(TM) are within the guidelines of the NCEP. Therefore the CardioCheck(TM) seems to be a nonreliable alternative to the conventional biochemical routine
dc.description.abstractYeni kullanılmaya başlayan CardioCheck(TM) taşınabilir lipid analizörünün, NCEP’e (National Cholesterol Education Program) göre değerlendirilmesi amaçlanmıştır. Total Kolesterol (TK), HDL Kolesterol (HDL-K) ve Trigliserid (TG) düzeyleri saptamak için, polikliniğe gelen yüz hasta çalışmaya dahil edilmiştir. Örnekler eş zamanlı olarak CardioCheck(TM) lipid analizöründe ve Biyokimya Laboratuvarında otomatik enzimatik metod ile çalışıldı. Tekrarlanabilirlik çalışması, normal ve yüksek kontrol serumunda yirmişer kez saptandı. Tekrarlanabilirlik çalışmasının varyasyon katsayısı (% CV) sonuçları; panel TK % 3.29 ( NCEP % ? 3) , panel HDL-K % 6.1 ( NCEP ? %6), panel TG % 18.9 ( NCEP ? % 5), TK % 5.3, HDL-K % 14.5, TG % 15.7 idi. CardioCheck(TM) tekrarlanabilirlik sonuçları NCEP’e göre değerlendirildi. CardioCheck(TM) lipid analizörünün geleneksel rutin ölçüme bir seçenek olacak kadar güvenilir olmadığı ortaya kondu
dc.identifier.dergiparkDergiPark: 610915
dc.identifier.endpage9
dc.identifier.issn1307-671X
dc.identifier.issue3
dc.identifier.startpage5
dc.identifier.urihttps://dergipark.org.tr/tr/download/article-file/793143
dc.identifier.urihttps://dergipark.org.tr/tr/pub/dtfd/issue/48257/610915
dc.identifier.urihttps://hdl.handle.net/20.500.12684/16155
dc.identifier.volume7
dc.language.isotr
dc.publisherDüzce Üniversitesi
dc.relation.ispartofDüzce Tıp Fakültesi Dergisi
dc.relation.publicationcategoryMakale - Ulusal Hakemli Dergi
dc.rightsinfo:eu-repo/semantics/openAccess
dc.snmzKA_20241228
dc.subjectSerum lipids
dc.subjectPoint of care test
dc.subjectLipid analyzer
dc.subjectSerum Lipidleri
dc.subjectHasta Başı Testler
dc.subjectLipid Analizörü
dc.titlePerformance of Cardiocheck (Tm) Portable Lipid Analyser
dc.title.alternativeTaşınabilir Lipid Analizörü Cardıocheck (Tm) Performansı