Comparison of the Lag Screw Placements for the Treatment of Stable and Unstable Intertrochanteric Femoral Fractures regarding Trabecular Bone Failure

dc.contributor.authorÇelik, Talip
dc.contributor.authorMutlu, İbrahim
dc.contributor.authorÖzkan, Arif
dc.contributor.authorKişioğlu, Yasin
dc.date.accessioned2020-04-30T13:32:14Z
dc.date.available2020-04-30T13:32:14Z
dc.date.issued2016
dc.departmentDÜ, Mühendislik Fakültesi, Biyomedikal Mühendisliği Bölümüen_US
dc.description.abstractBackground. In this study, the cut-out risk of Dynamic Hip Screw (DHS) was investigated in nine different positions of the lag screw for two fracture types by using Finite Element Analysis (FEA). Methods. Two types of fractures (31-A1.1 and A2.1 in AO classification) were generated in the femur model obtained from Computerized Tomography images. The DHS model was placed into the fractured femur model in nine different positions. Tip-Apex Distances were measured using SolidWorks. In FEA, the force applied to the femoral head was determined according to the maximum value being observed during walking. Results. The highest volume percentage exceeding the yield strength of trabecular bone was obtained in posterior-inferior region in both fracture types. The best placement region for the lag screw was found in the middle of both fracture types. There are compatible results between Tip-Apex Distances and the cut-out risk except for posterior-superior and superior region of 31-A2.1 fracture type. Conclusion. The position of the lag screw affects the risk of cut-out significantly. Also, Tip-Apex Distance is a good predictor of the cut-out risk. All in all, we can supposedly say that the density distribution of the trabecular bone is a more efficient factor compared to the positions of lag screw in the cut-out risk. © 2016 Talip Celik et al.en_US
dc.identifier.doi10.1155/2016/5470798en_US
dc.identifier.issn2314-5129
dc.identifier.scopusqualityN/Aen_US
dc.identifier.urihttps://dx.doi.org/10.1155/2016/5470798
dc.identifier.urihttps://hdl.handle.net/20.500.12684/181
dc.identifier.volume2016en_US
dc.indekslendigikaynakScopusen_US
dc.language.isoenen_US
dc.publisherHindawi Limiteden_US
dc.relation.ispartofJournal of Medical Engineeringen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.titleComparison of the Lag Screw Placements for the Treatment of Stable and Unstable Intertrochanteric Femoral Fractures regarding Trabecular Bone Failureen_US
dc.typeArticleen_US

Dosyalar

Orijinal paket
Listeleniyor 1 - 1 / 1
Yükleniyor...
Küçük Resim
İsim:
0181.pdf
Boyut:
2.92 MB
Biçim:
Adobe Portable Document Format
Açıklama:
Tam Metin / Full Text